Pocket Lab Essentials

Toolboxes / Pocket Lab Toolkit

Pocket Lab Toolkit: measured results and formulas

What the automatic colony count found on generated plates, where it is off, and the formulas every tool uses. The app shows the same tables on its Count screen. Check every count yourself: the automatic count is a start, corrected by a tap.

Automatic colony count on generated plates

Measured on generated plates, not on real photos: each 600 × 600 pixels of brightness only (0 to 255), a dish of radius 270 with a bright wall, agar at 100 and noise of ±8 levels, colonies drawn as discs 4 to 8 pixels in radius at a contrast of 50 unless the row says otherwise. Settings: sensitivity 50, light colonies, the default sizes, unless the row says otherwise.

Generated plates
PlateColoniesCountedError
Separate colonies, even light4040, each at its placenone
150 small colonies (radius 3 to 6)150150none
Empty plate (noise ±8; and ±16 at sensitivity 100)00none
Dark colonies, counted as dark4040none
Dark colonies, counted as light400all missed: set "Darker than the agar"
Light changing across the dish (135 levels side to side)4040none
Large colonies (radius 17 to 25, contrast 40 or 25), sensitivity 0, 50 and 1002525, each at its place, area within −15% to +20%none
Domed colonies (brightest in the middle, radius 6 to 12)4040none
Vignetting (10 to 40 levels darker at the wall), empty, light or dark, sensitivity 50 and 10000none
Vignetting, 40 dark colonies counted as dark4040none
A shadow 30 levels deep at the wall, light, sensitivity 0, 50 and 1004040none
The same, dark colonies counted as dark4040; 39 at sensitivity 100at 100 one colony by the wall joins the shadow and is left out as larger
A lamp's reflection, a 60 × 12 bar 150 levels bright, sensitivity 0 and 504040 or 41every colony found; the bar may count as one
The dish drawn 8% too big (the wall and table inside it)4040none: the wall is left out as larger
The dish cut off by the photo's edge, light or dark99none
Colonies inside the 5% margin at the wall40missed by design: the wall's edge is left out
Faint colonies (contrast 15) beside strong ones (contrast 50), sensitivity 020 + 20337 faint ones left out
The same at sensitivity 25 to 10020 + 2040none
Touching pairs (20 pairs, radius 6, 11 pixels apart)4040, each at its place, marked as splitnone
A clump of 5 packed close (radius 6, 7 pixels apart) among 20 alone2521packed so tight it has one peak: counts as one, not marked

Touching colonies are mostly missed, not made up: the more they overlap, the more are missed, about half when their centers are 0.6 of their radii apart. Measured on 4 generated plates a cell, each with 20 colonies alone and 10 groups of touching ones.

Touching colonies found
GroupCenters apartRadius 5 to 7Radius 4 to 8
20.9580 of 8080 of 80
20.8580 of 8072 of 80 (−10%)
20.7569 of 80 (−13.8%)61 of 80 (−23.8%)
20.656 of 80 (−30%)48 of 80 (−40%)
30.95119 of 120 (−0.8%)118 of 120 (−1.7%)
30.85116 of 120 (−3.3%)112 of 120 (−6.7%)
30.7599 of 120 (−17.5%)93 of 120 (−22.5%)
30.666 of 120 (−45%)64 of 120 (−46.7%)
50.95195 of 200 (−2.5%)199 of 200 (−0.5%)
50.85179 of 200 (−10.5%)169 of 200 (−15.5%)
50.75147 of 200 (−26.5%)135 of 200 (−32.5%)
50.691 of 200 (−54.5%)92 of 200 (−54%)

Not measured yet

Formulas

Each answer in the app shows its formula with your numbers and units.

ToolFormulaNotes
Molaritymass = concentration × volume × MWSolved for whichever of the four is left empty.
Molarityconcentration = mass ÷ (MW × volume)
Molarityvolume = mass ÷ (MW × concentration)
MolarityMW = mass ÷ (concentration × volume)
Dilution (C1V1 = C2V2)V1 = C2 × V2 ÷ C1Solved for whichever of the four is left empty; both concentrations must be the same kind (molar, mass per volume, % or ×).
Dilution (C1V1 = C2V2)V2 = C1 × V1 ÷ C2
Dilution (C1V1 = C2V2)C1 = C2 × V2 ÷ V1
Dilution (C1V1 = C2V2)C2 = C1 × V1 ÷ V2
Serial dilutiontransfer = V ÷ F; diluent = V − transferV is each tube's volume once it has its transfer; F the dilution at each step (10 for 1 in 10).
Percent solutionsgrams = % × milliliters ÷ 100w/v: grams per 100 mL.
Percent solutionssolute = % × final volume ÷ 100v/v.
Percent solutionssolute = % × total mass ÷ 100w/w.
Concentration conversionsper liter of solution: grams = moles × MW = % w/v × 10 = w/w fraction × ρ = v/v fraction × ρ(solute); N = M × equivalents per mole; molality = moles ÷ (ρ − grams) × 1000ρ is the solution's density in g/L; with none typed, 1 g/mL (a dilute water solution) is assumed and said. ppm and ppb are by mass. The equivalents per mole are typed, never guessed; v/v uses the pure solute's density.
Concentration conversionsstock volume = final strength × final volume ÷ stock strength (each as solute per liter of its own solution)A stock in any of the units above to a final solution in any other, each read with its own density.
Concentration conversionsSG = ρ ÷ ρ(water at the reference temperature)Water's density from Tanaka et al., Metrologia 38 (2001) 301, the formula the CIPM recommends (0 to 40 °C); against water at 4 °C unless another temperature is typed.
Concentration conversionsρ = SG × ρ(water at the reference temperature)
pH and buffers[H⁺] + [Na⁺] = [OH⁻] + C × Ka ÷ (Ka + [H⁺]), [OH⁻] = Kw ÷ [H⁺], pH = −log₁₀[H⁺]An acid's pH: the exact root of its charge balance with water's own ions (a strong acid has Ka ÷ (Ka + [H⁺]) = 1), never the √(Ka × C) shortcut, which is shown beside it. Kw is 10⁻¹⁴ (25 °C); concentrations stand for activities.
pH and buffers[H⁺] + C × [H⁺] ÷ (Ka + [H⁺]) = [OH⁻], [OH⁻] = Kw ÷ [H⁺], pH = −log₁₀[H⁺]A base's pH, read through the pKa of its conjugate acid as tables list it.
pH and bufferspH = pKa + log₁₀([base form] ÷ [acid form]) (Henderson–Hasselbalch); exactly, [base form] = C × (ν − k) + [OH⁻] − [H⁺]A buffer recipe, made exact by the charge balance: ν is the protons given up per acid molecule at the target pH, from the pKa typed and the same acid's steps below and above when typed; k is the protons the acid form has given up. Said when Henderson–Hasselbalch alone would miss by more than 0.05.
pH and buffersV(base) = V(acid) × (ν × C(acid) − [H⁺] + [OH⁻]) ÷ (C(base) + [H⁺] − [OH⁻])Strong base to bring an acid of up to three steps to a target pH: the charge balance after mixing.
pH and buffersd(pKa)/dT = −ΔrH° ÷ (R × T² × ln 10) (van 't Hoff, at 25 °C)The pKa list: pK at 25 °C and zero ionic strength, and ΔrH°, from Goldberg, Kishore and Lennen, J. Phys. Chem. Ref. Data 31 (2002) 231, Section 8 (NIST).
Gas lawsP = nRT ÷ VThe ideal gas law, solved for whichever of P, V, n and T is left empty. Worked in kPa, L, mol and K with R = 8.314 462 618 L·kPa/(mol·K) (N_A × k, exact since the 2019 SI). Close for most gases near room temperature and 1 atm; real gases depart from it at high pressure and near their boiling point.
Gas lawsV = nRT ÷ P
Gas lawsn = PV ÷ RT
Gas lawsT = PV ÷ nR
Gas lawsP1 = P2 × V2 × T1 ÷ (T2 × V1)The combined gas law, P1V1 ÷ T1 = P2V2 ÷ T2, for the same amount of gas, solved for whichever of the six is left empty; temperatures in kelvin.
Gas lawsV1 = P2 × V2 × T1 ÷ (T2 × P1)
Gas lawsT1 = T2 × P1 × V1 ÷ (P2 × V2)
Gas lawsP2 = P1 × V1 × T2 ÷ (T1 × V2)
Gas lawsV2 = P1 × V1 × T2 ÷ (T1 × P2)
Gas lawsT2 = T1 × P2 × V2 ÷ (P1 × V1)
Gas lawsp = x × P(total)Dalton's law of partial pressures; x is the gas's share by moles (for ideal gases, by volume too).
Gas lawsP(total) = p ÷ x
Gas lawsx = p ÷ P(total)
Gas lawsV = n × Vm, Vm = RT ÷ PAt IUPAC's STP (273.15 K, 100 kPa: Vm = 22.711 L/mol) or SATP (298.15 K, 100 kPa: 24.790 L/mol); the value at 1 atm (22.414 and 24.465 L/mol), as many textbooks use, is given beside it.
Gas lawsn = V ÷ Vm, Vm = RT ÷ P
Pressure and energyp in Pa = value × (Pa per unit)Exact definitions: 1 atm = 101 325 Pa, 1 bar = 10⁵ Pa, 1 Torr = 1/760 atm, 1 mmHg = 133.322 387 415 Pa (conventional), 1 psi = 6894.757 293 168 Pa. Absolute pressures.
Pressure and energyE in J = value × (J per unit)Exact definitions: 1 cal = 4.184 J (thermochemical), 1 kWh = 3.6 × 10⁶ J, 1 eV = 1.602 176 634 × 10⁻¹⁹ J, 1 L·atm = 101.325 J.
Centrifuge (RCF and rpm)RCF = 1.118 × 10⁻⁵ × r × rpm²r in cm. The constant is (2π/60)² ÷ 980.665 cm/s², to four figures as rotor sheets print it.
Centrifuge (RCF and rpm)rpm = √(RCF ÷ (1.118 × 10⁻⁵ × r))
Temperature°F = °C × 9/5 + 32; K = °C + 273.15
Temperature°C = (°F − 32) × 5/9; K = °C + 273.15
Temperature°C = K − 273.15; °F = °C × 9/5 + 32
Mass and molar concentrationmass concentration = molar concentration × MW
Mass and molar concentrationmolar concentration = mass concentration ÷ MW
Cells to seedvolume = cells wanted ÷ cell density
Master mixtotal = volume per reaction × reactions × (1 + overage ÷ 100)
Molar massM = sum of (atoms × standard atomic weight)Standard atomic weights: CIAAW, Abridged Standard Atomic Weights 2024 (ciaaw.org).
Reaction (limiting reagent and yield)n = mass × purity ÷ 100 ÷ M; limiting reagent: smallest n ÷ coefficient; each amount = coefficient × n(limiting) ÷ coefficient(limiting)M from the formula (Molar mass above); purity 100% when none is typed, and an amount typed in moles is taken as pure. The equation is taken as typed, never balanced for you; equivalents are moles ÷ moles of the limiting reagent.
Reaction (limiting reagent and yield)n = mass × purity ÷ 100 ÷ M; limiting reagent: smallest n ÷ coefficient; each amount = coefficient × n(limiting) ÷ coefficient(limiting); % yield = actual ÷ theoretical × 100With the amount of a product made typed: its percent yield, by moles (the same ratio by mass).
Ligationinsert mass = vector mass × (insert length ÷ vector length) × ratioThe ratio is insert to vector by moles (3 for 3:1), as in Promega's pGEM-T Vector Systems technical manual (TM042). Amounts in fmol use 650 g/mol per base pair, an average.
DNA/RNA (A260)concentration = A260 × 50 µg/mL × dilution ÷ path (cm)Double-stranded DNA: 50 µg/mL per A260 unit at 1 cm; RNA 40, single-stranded DNA 33 (Sambrook and Russell, Molecular Cloning, 3rd ed.). Purity: A260/A280 about 1.8 for DNA and 2.0 for RNA; A260/A230 about 2.0 to 2.2.
DNA/RNA (A260)concentration = A260 × 40 µg/mL × dilution ÷ path (cm)RNA.
DNA/RNA (A260)concentration = A260 × 33 µg/mL × dilution ÷ path (cm)Single-stranded DNA.
DNA/RNA (A260)concentration = A260 × dilution ÷ (ε260 × path)An oligo, by the Beer-Lambert law with its own ε260 in L/(mol·cm).
Protein (A280)concentration = A280 × dilution ÷ (ε280 × path) × MWThe Beer-Lambert law with the protein's own ε280 in L/(mol·cm) and molecular weight.
Protein (A280)concentration = A280 × 1 mg/mL × dilution ÷ path (cm)A rough shortcut: real proteins differ several-fold.
OD600 and growthcells/mL = OD600 × dilution × cells per mL at OD 18 × 10⁸ cells/mL at OD 1 for E. coli when none is typed, an approximate figure.
OD600 and growthdoubling time = t × ln 2 ÷ ln(N2 ÷ N1)Exponential growth between two readings.
Seedingcells = cells per cm² × area per well × wells; volume = cells ÷ cells per mL
Seedingsuspension per vessel = volume ÷ split ratioSplit 1:N.
Colonies per mL (CFU)CFU/mL = colonies × dilution factor ÷ volume plated (mL)Reported to two significant figures, rounded as FDA BAM ch. 3 rounds. Outside 15 to 300 colonies per plate the result is marked as an estimate; with no colonies it is less than the detection limit (1 colony at that dilution).
Hemocytometercells/mL = mean cells per large square × dilution factor × 10⁴For 1 mm² squares 0.1 mm deep: 10⁴ is 1 mL over one square's volume (0.1 µL). A 0.2 mm chamber uses 5000.
Hemocytometerviability = live cells ÷ (live + dead cells) × 100
Differential countpercent = count of the kind ÷ total counted × 100Only keys in the total count toward it; a key left out of the total is said per 100 counted.
Differential count95% range = (p + z²/2N ± z·√(p(1 − p)/N + z²/4N²)) ÷ (1 + z²/N), z = 1.96Wilson score interval (Wilson 1927), as recommended by Newcombe (1998); p = count ÷ N, N = total counted.
Differential countabsolute count = count of the kind ÷ total counted × total count typedIn the unit of the total typed.
Punnett squareshare = squares in the class ÷ all squaresMendel’s segregation and independent assortment: each gamete equally likely, genes taken to be unlinked.
Chi-squareE = N × part ÷ sum of parts
Chi-squareχ² = Σ (O − E)² ÷ EPearson’s chi-square; also used by Hardy-Weinberg and Map distance.
Chi-squaredegrees of freedom = classes − 1
Chi-squarep = Q(df ÷ 2, χ² ÷ 2)Q is the regularized upper incomplete gamma, by its series and continued fraction (Numerical Recipes, section 6.2). A warning shows when an expected count is under 5 (Cochran 1954).
Hardy-Weinbergp = (2 × AA + Aa) ÷ (2 × N)q = 1 − p.
Hardy-Weinbergexpected AA = p² × N, Aa = 2pq × N, aa = q² × N
Hardy-Weinbergdegrees of freedom = genotypes − 1 − 1One less for the allele frequency, estimated from the same counts.
Mouse crossshare = wanted pups ÷ equally likely pupsEvery way the parents’ gametes can meet, each allele and each sex with chance 1/2; loci taken to be unlinked.
Mouse crosswanted per litter = share × litter size × survivalLitter size and survival are your own numbers.
Mouse crosslitters = N ÷ wanted per litter, rounded upAn average.
Mouse crosspairs = litters ÷ litters per pair, rounded up
Mouse crossn = litters × litter size × survival, rounded down
Mouse crossP(at least N) = Σ C(n, k) share^k (1 − share)^(n − k), k = N to nThe binomial upper tail.
Map distanceRF = recombinants ÷ totalFrom a two-point test cross.
Map distancemap distance = RF × 100 cM1 map unit (cM) = 1% recombination.
Map distanceHaldane d = −½ ln(1 − 2RF) × 100 cMHaldane 1919: no interference. Given for RF under 0.5.
Map distanceKosambi d = ¼ ln((1 + 2RF) ÷ (1 − 2RF)) × 100 cMKosambi 1944: some interference. Given for RF under 0.5.
Decay and countsA = A₀ × 2^(−t ÷ T½)The same as A₀ × e^(−λt) with λ = ln 2 ÷ T½. Half-lives: IAEA Nuclear Data Section, LiveChart of Nuclides (ENSDF evaluated data), read 2026-11-19.
Decay and countst = T½ × log₂(A₀ ÷ A)The time for the activity to fall from A₀ to A.
Decay and countsT½ = t × ln 2 ÷ ln(A₀ ÷ A)From two readings t apart. A small fall between them makes it uncertain.
Decay and countsnet rate = N_g ÷ t_g − N_b ÷ t_b, σ = √(N_g ÷ t_g² + N_b ÷ t_b²)Poisson counting statistics (Knoll, Radiation Detection and Measurement, chapter 3).
Decay and countsactivity (Bq) = net counts per second ÷ (efficiency × emission yield)1 Ci = 3.7 × 10¹⁰ Bq; 1 dpm = 1/60 Bq.
Decay and countsN = (z ÷ r)²Counts for a relative uncertainty r: z = 1 for one standard deviation, 1.96 for 95%. Poisson counting statistics (Knoll, Radiation Detection and Measurement, chapter 3).
Decay and countsT = (√R_g + √R_b)² ÷ (R_n² × (r ÷ z)²), t_g ÷ t_b = √(R_g ÷ R_b)The shortest total counting time for a net rate known to r, and its split between sample and background. Poisson counting statistics (Knoll, Radiation Detection and Measurement, chapter 3).
Decay and countsL_C = 1.645 × √(2B), L_D = 1.645² + 2 × L_C (≈ 2.71 + 4.65 × √B)Currie (1968), Analytical Chemistry 40, 586: a blank counted as long as the sample, 5% false positives and false negatives.
Decay and countsMDA (Bq) = L_D ÷ (efficiency × emission yield × counting time in s)
Pipette CheckV = m × Z, Z = 1 ÷ (ρ_w − ρ_a) × (1 − ρ_a ÷ ρ_b)Gravimetric volume (as in EURAMET cg-19): m the weighing, ρ_w water, ρ_a air, ρ_b the balance's weights (8.0 g/mL).
Pipette Checkρ_w = a5 × (1 − (t + a1)² × (t + a2) ÷ (a3 × (t + a4)))Air-free water, Tanaka et al. (2001), Metrologia 38, 301: a1 = −3.983035 °C, a2 = 301.797 °C, a3 = 522528.9 °C², a4 = 69.34881 °C, a5 = 999.97495 kg/m³.
Pipette Checkρ_a = (0.34848 × p − 0.009 × h × e^(0.061 × t)) ÷ (273.15 + t)Moist air in kg/m³, p in hPa, h in % relative humidity, t in °C (CIPM-2007 approximation, as in OIML R 111-1).
Pipette Checksystematic error = mean volume − nominal volume; random error = standard deviation (n − 1); CV = SD ÷ mean × 100Passed or failed only against the limits the user types.